ProChek® ID Process Indicator Strip, Steam and Chemical Vapor, 4" x 3/4", 250/bx, 6 bx/cs SM4340

Item Number: SM4340
Packaging: bx
Manufacturer: Certol International
Rx
Grouped product items
Otions Quantity Price
Case
6 Box/Case
$141.49
Quick View
Box
250/Box
$23.69
Quick View

Features & Benefits:

  • The ProChek® ID Process Indicator is an economical single parameter chemical indicator strip and a useful component of a comprehensive sterilization quality assurance program.
  • Economical Process Indicator:
    • Place a strip inside every pouch, pack or cassette prior to steam sterilization.
    • Indicators turn color from white to black in steam process or white to brown in chemical vapor process.
    • Visible assurance that package or load was exposed to the sterilization process.
  • All Pouches and Packs Should Include Internal and External Chemical Indicators:
    • Early indicator of possible sterilizer malfunction or operator error.
    • Provides feedback on sterilization functions between weekly biological monitoring (spore tests).
  • Meets CDC Guidelines for an Internal Chemical Indicator:
  • "Internal chemical indicators should be used inside each package to ensure the sterilizing agent has penetrated the packaging material."
  • Economical single parameter indicator - include a strip in every pouch.
  • Temperature sensitive indicator ink turns from white to black in steam process; from white to brown in chemical vapor.
  • Dimensions: 4" x 3/4".
  • Parameters: Measures temperature.
  • Indicator Ink: Steam and chemical vapor specific ink.

Application:

  • When used as directed ProChek® ID process indicator strips give visible confirmation that the contents of a pouch or pack have been exposed to the steam or chemical vapor sterilization process.
  • ProChek® ID is recommended to be used inside every pouch, wrapped cassette pack or surgery pack. It may also be used in open, unwrapped loads.
Show More

FAQs

  • I use a steam autoclave sterilizer in my office. How do I know if it is working correctly?

    For sterilization quality assurance, CDC Guidelines advise the use of three components:

    • Physical monitors such as temperature, pressure and length of cycle using gauges on sterilizer. New sterilizers feature print-outs for each cycle.
    • Chemical ink monitors (external and internal) for every package and every load.
    • Biological monitors (spore tests) weekly.
  • How do ProChek products fit into a quality assurance program?

    ProChek ID Indicator Tape (PIT100, PIT120, PIT340) is an external Class 1 process indicator. It reacts to identify packages and pouches exposed to the sterilization process. The tape may be used to close and seal packages and/or short strips may be applied externally. ProChek ID Indicator Strips (SM4340) are internal Class 1 indicators reacting to steam moisture and heat. The ink changes from white to dark brown/black in steam or medium brown in chemical vapor. This product meets the CDC and AAMI ST79 guidelines for an internal chemical indicator to be used inside pouches, cassettes and packs to verify penetration of heat and steam inside the package. ProChek S Steam Sterilization Integrators (TSI440) are Class 4 multi-parameter internal chemical monitors. The term "integrator" means the special chemical ink changes completely from yellow to blue/purple as it reacts to steam and heat over a time period that is near the parameters required to kill spores. This level of internal chemical monitor provides a higher level of assurance for surgery packs and instrument cassettes.

  • Do I need to use both external and internal indicators with every pouch and package?

    CDC and AAMI ST79 Guidelines specify that if the internal indicator cannot be seen from within the unopened package, then separate external indicators should be applied to the pouch or package. For speed and safety most facilities find it helpful to use both for every package.

  • Can ProChek S be used for all types of table top sterilizers?

    It is specifically engineered for use with standard steam sterilizers (gravity and pre-vac). It is not suitable for monitoring chemical vapor "Chemiclaves", dry heat or STATIM® pulse sterilizers.

  • Do chemical monitors such as ProChek S take the place of biological spore tests?

    No. Your sterilization monitoring program must include biological spore tests, preferably at least once a week in each operational sterilizer in use.

  • I use sterilization pouches with "built-in" internal indicators. Do I still need separate internal monitoring strips?

    Most self-seal sterilization pouch brands, including Certol ProView® plus pouches, now feature built-in external and "internal" indicators. Technically this may be adequate for small pouches with a few instruments. But the "built-in" indicator cannot monitor adequate steam penetration inside instrument cassettes or thick surgery packs that are processed inside selfseal pouches. Therefore it is recommended that a separate internal monitoring strip be placed inside the cassette or pack prior to packaging and sterilization. ProChek ID strips (SM4340) may be used but ProChek S multiparameter Class 4 integrators (TSI440) provide a higher level of assurance for this application.

  • Can I use the ProChek ID Indicator Tape in my dry heat sterilizer or Chemiclave®?

    No, the tape is specific to the steam process. It will become brittle and char in the dry heat process. The indicator requires both steam and heat to change color and will not properly change (darken) in the low moisture environment of the Chemiclave®.

  • What is the protocol to follow when chemical indicators do not change color or in the case of ProChek S, do not turn completely blue/purple?

    Do not release the load of instruments for use. Check the physical controls (temperature, cycle time, etc.) and print out if available. Verify that correct cycle time, temperature and pressure are in use. Determine if the sterilizer is overloaded. Evaluate how items are loaded. Packages placed paper to paper or stacked flat instead of on edge may not allow adequate circulation and penetration of steam. Check the condition of the sterilizer door gasket. Ensure adequate level of distilled water or chemical vapor solution. Review error codes displayed on models with this feature. Repackage and reprocess all items with fresh chemical indicators. Re-evaluate after processing. If all chemical monitors continue to fail, contact equipment repair person. Check records for date of last successful biological spore test and date of most recent passing chemical indicators. Meanwhile, keep sterilizer out of service. Recall loads of sterilized items where process monitor results are not known or not passing.